A comparative field evaluation of six medicine quality screening devices in Laos

PLOS Neglected Tropical Diseases
Published
30 Sep 2021
Authors
Céline Caillet, Serena Vickers, Stephen Zambrzycki, Facundo M. Fernández, Vayouly Vidhamaly, Kem Boutsamay, Phonepasith Boupha, Pimnara Peerawaranun, Mavuto Mukaka, Paul N. Newton

Abstract

Background

Medicine quality screening devices hold great promise for post-market surveillance (PMS). However, there is little independent evidence on their field utility and usability to inform policy decisions. This pilot study in the Lao PDR tested six devices’ utility and usability in detecting substandard and falsified (SF) medicines.

Methodology/principal findings

Observational time and motion studies of the inspections by 16 Lao medicine inspectors of 1) the stock of an Evaluation Pharmacy (EP), constructed to resemble a Lao pharmacy, and 2) a sample set of medicines (SSM); were conducted without and with six devices: four handheld spectrometers (two near infrared: MicroPHAZIR RX, NIR-S-G1 & two Raman: Progeny, Truscan RM); one portable mid-infrared spectrometer (4500a), and single-use paper analytical devices (PAD). User experiences were documented by interviews and focus group discussions.

Significantly more samples were wrongly categorised as pass/fail with the PAD compared to the other devices in EP inspections (p<0.05). The numbers of samples wrongly classified in EP inspections were significantly lower than in initial visual inspections without devices for 3/6 devices (NIR-S-G1, MicroPHAZIR RX, 4500a). The NIR-S-G1 had the fastest testing time per sample (median 93.5 sec, p<0.001). The time spent on EP visual inspection was significantly shorter when using a device than for inspections without devices, except with the 4500a, risking missing visual clues of samples being SF. The main user errors were the selection of wrong spectrometer reference libraries and wrong user interpretation of PAD results. Limitations included repeated inspections of the EP by the same inspectors with different devices and the small sample size of SF medicines.

Conclusions/significance

This pilot study suggests policy makers wishing to implement portable screening devices in PMS should be aware that overconfidence in devices may cause harm by reducing inspectors’ investment in visual inspection. It also provides insight into the advantages/limitations of diverse screening devices in the hands of end-users.

Author summary

Substandard and falsified (SF) medicines threaten the lives of millions of people, especially where pharmaceutical legislation and regulation are limited. Screening for SF medicines in supply chains (‘post-market surveillance’) by medicine inspectors is crucial but currently relies only on subjective visual inspection of medicines in most countries. Many innovative portable screening technologies now exist and could be key additional assets to the current practice, but none have been extensively evaluated for medicine quality post-market surveillance. We assessed the utility and usability of six screening devices in the hands of Lao medicines inspectors in a pharmacy constructed to resemble a Lao pharmacy. Five spectrometers showed promising accuracies to identify falsified medicines, but difficulties to correctly set them up before running tests were observed. Reading and interpreting colour barcodes of the ‘paper analytical cards’–a lab-on-a-chip test by inspectors were difficult, leading to lower accuracy than with spectrometers. The study suggests that overconfidence in devices may cause harm by reducing inspectors’ investment in visual inspection—a crucial step to identify falsified medicines. Advantages/limitations of the devices are also documented to inform policy.